Difference between revisions of "Minutes-7-10-2014"

From GlueXWiki
Jump to: navigation, search
m (Text replacement - "/halldweb1.jlab.org/" to "/halldweb.jlab.org/")
 
(24 intermediate revisions by one other user not shown)
Line 1: Line 1:
July 3, 2014 FDC+CDC meeting
+
July 10, 2014 FDC+CDC meeting
  
 
= Agenda =
 
= Agenda =
 
# Gas system
 
# Gas system
# DAQ (Beni)
+
# DAQ (Beni)
# CDC status (Mike, Beni)
+
# CDC status [http://www.jlab.org/Hall-D/detector/fdc/CDC/CDC_noise_map.png CDC noise studies] (Mike, Beni)
# FDC status [http://www.jlab.org/Hall-D/detector/fdc/DAQ/DAQ_scope_checks_July_August_2014.xlsx] (Lubomir)
+
# FDC status [http://www.jlab.org/Hall-D/detector/fdc/DAQ/fADC_raw_data0_2.pdf CDC raw data], [http://www.jlab.org/Hall-D/detector/fdc/DAQ/fADC_raw_data4_4.pdf FDC raw data], [http://www.jlab.org/Hall-D/detector/fdc/DAQ/DAQ_scope_checks_July_August_2014.xlsx] (Lubomir)
#* [http://www.jlab.org/Hall-D/detector/fdc/DAQ/FDC_Data.ods noise studies] (Luke)
+
#* [http://www.jlab.org/Hall-D/detector/fdc/commissioning/FDC_Data.ods FDC noise studies], [http://www.jlab.org/Hall-D/detector/fdc/commissioning/200Hznoise.JPG] (Luke)
 
# Electronics (Chris, Nick)
 
# Electronics (Chris, Nick)
 
# Engineering (Bill)
 
# Engineering (Bill)
 
# Other
 
# Other
  
[https://halldweb1.jlab.org/wiki/index.php/Connect_to_Bluejeans_Meetings Instructions for Bluejeans meeting connection]
+
[https://halldweb.jlab.org/wiki/index.php/Connect_to_Bluejeans_Meetings Instructions for Bluejeans meeting connection]
  
 
FDC meeting ID: 290664653
 
FDC meeting ID: 290664653
 
<!--
 
  
 
= Minutes =
 
= Minutes =
  
Participants: Naomi (CMU), Beni, Dave, Chris, Mike, Simon, Fernando, Eugene, Glenn, and Lubomir (JLab).
+
Participants: Curtis, Naomi (CMU), Luke, Eugene, Dave, Chris, Nick, Simon, Mike, Beni, Fernando, and Lubomir (JLab).
 
+
== Readiness review ==
+
 
+
- Drafts (very preliminary) for the CDC and FDC commissioning were prepared for the readiness review (page linked above). For the FDC the plan so far is to use two gas mixtures, 90/10 and 40/60, the first one is better for the alignment. The target position(s) and turning on/off the magnetic field needs to be coordinated. 
+
  
 
== Gas system ==
 
== Gas system ==
  
- Last Friday, at ~8am, the power in the gas room was turned off for ~15 min due to a planned electrical work there. After turning it ON, the setting in the gas control panel were wrong (40mA, while the range is 4-20mA). As a result: (1) the flow was doubled and (2) the PLS didn't turned OFF the flow controllers when the mixing tanks reached 12psi. Soon the pressure reached 15psi and the two relief valves were continuously engaged for some time.  
+
- Last Thursday after the FDC meeting Dave and Lubomir connected the power of the gas controls in the gas room to a UPS. When the power was restored the 4-20mA setting of the two masters were different (40mA as usual), but also the settings of one slave channel did change, that never happened before. Dave will see if he can read the settings and either restore them correctly, or issue an alarm. This has to be discussed also with Hovanes.
  
- After the meeting we installed an UPS (in preparation for the holiday weekend) there. Dave will contact the company to find a solution, at least to change the default settings to what we need.
+
- On Saturday the switch on the CO2 bottles didn't work when one set of bottles got empty; it was supposed to switch to the full set of bottles. For one-two hours there was no CO2 flow into the FDC mixing tank, and 0.2l/m (instead of 0.9) was going into the CDC mixing tank. The bottom half of the CDC and 25% of the FDC channels tripped. Dave can implement an alarm, but everybody agrees some kind of a messaging system is needed, as well. Scott contacted the gas company to replace the switch. This switch was working before, and we have the same switch for the Argon bottles.
  
 
== DAQ ==
 
== DAQ ==
  
- Beni: can read now all the 18 crates (CDC+FDC). Sometimes DAQ crashes fatally, so that you have to reboot the computer (suspect InfiniBand). Also somehow the bios on gluon51 failed; will be replaced.  
+
- Beni: The attempt to read synchronously the BCAL and the CDC failed. It's not clear why, the ROCs where receiving the data but then they got lost somewhere. Will try it again tomorrow.  
  
- Beni made independent initialization for the FDC and CDC, now they can have different pulse windows.  
+
- Beni: there's a problems with the relation between the pedestal positions and the DAC values in the fADC. The relation is different for the CDC and the FDC. For the CDC the pedestals are at ~500 and all the DAC values are set to 0x9000 (hex), and for the FDC the pedestals are at ~1000 but the DAC values are set all at 0x8000 (see raw data plots linked to item 4). Fernando explained this by the difference in the pre-amp gains (factor of 4).
  
== CDC noise ==
+
== CDC update ==
 
+
- Fernando suspects the noise may come from the CDC itself, since it depends on the HV, and suggested increasing the CO2 content while keeping the same HV; the idea is to exclude the HV modules as a source of the noise. Eugene/Naomi: same effect will be if decrease/increase the HV by few hundred volts. Also it's difficult to explain with gas discharges the correlation we see in the noise (see plots from the last meeting). In any case, first Mike will study systematically where on the CDC we see the noise.
+
 
+
- Lubomir looked with a scope on the FDC and found noise with very similar pattern on several special channels. That was on the most upstream cathodes in the first and second package, but not in the other two, and always in cards that have both long and short strips (more susceptible to noise). It may indicate that the noise comes from the CDC or the mesh but one has to study this more systematically.
+
  
 +
- Mike studied the noise in the CDC using raw fADC data. He applies a negative threshold (below the pedestal) and plots all the channels that pass the threshold as function of the sample number: there's clear correlations between the channels. The positions of these noisy channels is shown in the plot attached: mostly at the 6-9 o'clock sector when looking downstream.
 +
 
 
== FDC update ==
 
== FDC update ==
  
- Lubomir started analyzing the fADC data from the last DAQ runs. The pedestal widths for the CDC are about a factor of 2 bigger than for the FDC: 20, 10 fADC units respectively. The widths in the FDC depend on the size of the strips.
+
- Lubomir looked at the raw fADC data: plots linked above represent the distributions of the ADC values (from all the samples) as function of the channel. The plots are very useful since one can identify different kinds of problems: noisy channels when the area below the pedestal is populated, missing bit (like in the CDC plot ch.63), detector problem when the signals are small (like chan.71 in the FDC plot), connection issue when there's only a thin pedestal. Another problem is like ch.69 FDC plot: there's no pedestal visible, all values are overflown, however, if you look with a scope with differential probe we do see normal signals. If you connect another "good" cable to the same fADC connector everything is OK. What is then the difference between the fADC and the differential probe? Is this problem related to the pedestal problem Beni discussed above?
  
- Important is that FDC sees many events with the CDC trigger; the explanation: most of the cosmic events are part of showers. The timing of the first maximum resembles the standard FDC drift time spectrum (histogram linked above). Beni will use this histogram to adjust the FDC pulse window from 20th to 100th sample, thus cutting the size of the files by 2. Before it was 0 to 160, as in the CDC.
+
- Summary of these observation are shown in a spreadsheet linked above (only half of the crates are included so far). The number of bad detector channels is about the same as we had right after the installation as studied with a scope.
  
== Electronics ==
+
- Luke showed his studies (linked) of the noise in the FDC using a scope. He was looking for similar correlated noise that is seen in the CDC. This kind of noise is well visible only on these long strip that have short strips next to them (due to having ground from both sides of the frame). Luke's summary: most of the noise in the first and second package and mostly in the outer cells (one and six), frequencies 200kHz, 80kHz and 40kHz, amplitudes 10-15mV. Interestingly, if the cable is not well plugged into the adapter card, one can see the same noise on all the channels and the noise increases when the card is close to the VME crates (see picture).
 
+
- Fernando: two CAEN modules back from repair. Chris/Nick will test them before using in the Hall.
+
  
 
== Engineering ==
 
== Engineering ==
  
- Eugene: on the readiness review there was a remark that people stopped using plastic tubes for Fluorinert, only metal. Tim answered that the tube material was chosen to work for Fluorinert.
+
- Bill (not at the meeting): the microscope holder for the cathode scan is ready. Need to connect the motor the stage and to wire the motor controller and the stoppers (Dave). Luke will work on the scanning project.
 
+
- Dave: put new mark on the Fluorunert level. Moved all the hardware needed for the cathode scan, the motor, controller and computer, in the ESB.
+
 
+
== Other ==
+
 
+
- Luke Robison, a graduate student from Northwestern University will work on FDC projects for ~2 months: noise studies, cathode scan.
+
 
+
-->
+

Latest revision as of 05:58, 1 April 2015

July 10, 2014 FDC+CDC meeting

Agenda

  1. Gas system
  2. DAQ (Beni)
  3. CDC status CDC noise studies (Mike, Beni)
  4. FDC status CDC raw data, FDC raw data, [1] (Lubomir)
  5. Electronics (Chris, Nick)
  6. Engineering (Bill)
  7. Other

Instructions for Bluejeans meeting connection

FDC meeting ID: 290664653

Minutes

Participants: Curtis, Naomi (CMU), Luke, Eugene, Dave, Chris, Nick, Simon, Mike, Beni, Fernando, and Lubomir (JLab).

Gas system

- Last Thursday after the FDC meeting Dave and Lubomir connected the power of the gas controls in the gas room to a UPS. When the power was restored the 4-20mA setting of the two masters were different (40mA as usual), but also the settings of one slave channel did change, that never happened before. Dave will see if he can read the settings and either restore them correctly, or issue an alarm. This has to be discussed also with Hovanes.

- On Saturday the switch on the CO2 bottles didn't work when one set of bottles got empty; it was supposed to switch to the full set of bottles. For one-two hours there was no CO2 flow into the FDC mixing tank, and 0.2l/m (instead of 0.9) was going into the CDC mixing tank. The bottom half of the CDC and 25% of the FDC channels tripped. Dave can implement an alarm, but everybody agrees some kind of a messaging system is needed, as well. Scott contacted the gas company to replace the switch. This switch was working before, and we have the same switch for the Argon bottles.

DAQ

- Beni: The attempt to read synchronously the BCAL and the CDC failed. It's not clear why, the ROCs where receiving the data but then they got lost somewhere. Will try it again tomorrow.

- Beni: there's a problems with the relation between the pedestal positions and the DAC values in the fADC. The relation is different for the CDC and the FDC. For the CDC the pedestals are at ~500 and all the DAC values are set to 0x9000 (hex), and for the FDC the pedestals are at ~1000 but the DAC values are set all at 0x8000 (see raw data plots linked to item 4). Fernando explained this by the difference in the pre-amp gains (factor of 4).

CDC update

- Mike studied the noise in the CDC using raw fADC data. He applies a negative threshold (below the pedestal) and plots all the channels that pass the threshold as function of the sample number: there's clear correlations between the channels. The positions of these noisy channels is shown in the plot attached: mostly at the 6-9 o'clock sector when looking downstream.

FDC update

- Lubomir looked at the raw fADC data: plots linked above represent the distributions of the ADC values (from all the samples) as function of the channel. The plots are very useful since one can identify different kinds of problems: noisy channels when the area below the pedestal is populated, missing bit (like in the CDC plot ch.63), detector problem when the signals are small (like chan.71 in the FDC plot), connection issue when there's only a thin pedestal. Another problem is like ch.69 FDC plot: there's no pedestal visible, all values are overflown, however, if you look with a scope with differential probe we do see normal signals. If you connect another "good" cable to the same fADC connector everything is OK. What is then the difference between the fADC and the differential probe? Is this problem related to the pedestal problem Beni discussed above?

- Summary of these observation are shown in a spreadsheet linked above (only half of the crates are included so far). The number of bad detector channels is about the same as we had right after the installation as studied with a scope.

- Luke showed his studies (linked) of the noise in the FDC using a scope. He was looking for similar correlated noise that is seen in the CDC. This kind of noise is well visible only on these long strip that have short strips next to them (due to having ground from both sides of the frame). Luke's summary: most of the noise in the first and second package and mostly in the outer cells (one and six), frequencies 200kHz, 80kHz and 40kHz, amplitudes 10-15mV. Interestingly, if the cable is not well plugged into the adapter card, one can see the same noise on all the channels and the noise increases when the card is close to the VME crates (see picture).

Engineering

- Bill (not at the meeting): the microscope holder for the cathode scan is ready. Need to connect the motor the stage and to wire the motor controller and the stoppers (Dave). Luke will work on the scanning project.