Glue light guide to the PMT

From GlueXWiki
Revision as of 12:50, 22 February 2022 by Somov (Talk | contribs) (Created page with "gallery> File:glue_lg1.jpg | Gluing fixture File:glue_lg2.jpg | DYMAX glue loaded to the syringe File:glue_lg3.jpg | DYMAX BlueWave curing tool File:glue_lg4.jpg | Cure UV...")

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

gallery> File:glue_lg1.jpg | Gluing fixture File:glue_lg2.jpg | DYMAX glue loaded to the syringe File:glue_lg3.jpg | DYMAX BlueWave curing tool File:glue_lg4.jpg | Cure UV glue from the top of the LG File:glue_lg5.jpg | Cure UV glue from the sides File:glue_lg6.jpg | Light guide glued to the PMT

</gallery> Warning: UV protection glasses (see Fig. 3) have to be used during UV gluing prcedure

  • Clean the PMT face with isopropanol and insert it inside the gluing fixture, slide the PMT up from the bootom
  • Make sure that DYMAX UV glue is loaded to the syringe
  • Apply one drop of the glue on the PMT face
  • Make sure that the light guide surface is clean
  • Slide the light guide from the top and put in contact with the PMT. Make sure that there are no air bubbles in the glue between the PMT and LG
  • Switch on UV gluing tool. Cure UV glue from the top and two sided. The typical light exposure time is 6 sec for each side
  • Slide the PMT with LG to the bottom of the fixture