RunPeriod-2017-01 Validation

From GlueXWiki
Jump to: navigation, search

Problems/Features

  • Note: Runs 30796-30807 have problematic FCAL gain settings which is not well modeled by simulation. Until the simulation is updated, we will take these runs off of the approved list (1 June 2021).

Sean

  • Known trigger changes: Run 30314, Run 30320 - Spring 2017 Trigger Summary
  • BCAL occupancy problems around 30401, fixed soon after by pedestal recalibration
  • TDC problems
    • 32ns shifts still happening [e.g. approx 1/6 of runs in BCAL]
    • individual chips [groups of 4 channels] and individual chips shifting by multiples of 4 ns.
  • TOF ADC needs to be recalibrated 30274 - 30389
  • Problem TAGH runs: 30286, 30320, 30322, 30323, 30324, 30333, 30334, 30335, 30355, 30459, 30465, 30830, 30833, 30834, 30835, 30838, 30841, 30842, 30844, 30847, 30848, 30855, 30857, 30858, 30888, 30893, 30896, 30902, 30903, 30923, 30924, 30926, 30927, 30929, 30954, 30955, 30957, 30962, 30963, 30995, 31001, 31002, 31018
  • Known tagger problems: 30330, 30336, 30355, 30463, 30473(32ns), 30826(32ns), 30830, 30838, 30842, 30844, 30847, 30848, 30855, 30857, 30888, 30889, 30895, 30898, 30899, 30920, 30928, 30947, 30951, 30956, 30958
    • low-rate runs
      • TAGM: 30666, 30796-30835, 30841, 30410(weird), 30420(bad bunches), 30428(base time), 30490(bad bunches), 30587(bad bunches), 30593(bad bunches), 30596(bad bunches), 30674(base time), 30701(base time)
      • TAGH: 30424(offset), 30460(base time), 30630(base time)
  • Second pass: 3/27/17
    • Missing TAGH channels: 30330
    • TAGM calibration problems: 30596, 30450
    • Empty jobs: 30405, 30346
    • TAGH base offset: 30465

Note that very low rate runs are: 30980-30982, 31050-31057

  • Third pass: 3/29/17
    • Follow up: 30771, 30786, 30804, 30808
  • Current status: 5/5/17
    • TAGM
      • Column 26,48 are very poor quality, difficult to calibrate
      • Various other problems, some are within window, some aren't. Need another launch...

Mahmoud

I looked at the monitoring plots and compared them to run30441. The changes from the expected plots are as follows:

HLDT PID System Timing Plots


Run Number comments
30786 7-10 times less statistics
30771 7-10 times less statistics & sector 13 Δt shifted down to -4 ns
30701 Sector 13 Δt shifted down to -8 ns
30682, 30449 TOF Δt second plane peak is split.
30667,30652,30635/4,30618,30593,30489, 30471,30436,30437,30404/3/2/1 A weaker second lin appears at 6 ns for several SC sectors
30476 Sectors 11,13, and 26 Δt shifted down to -4 ns
30389--->30343 TOF Δt peaks seems significantly different than the ideal plot
30337---->30333 4-5 times less statistics and TOF Δt peaks seems significantly different than the ideal plot.
30332---->30327 A weaker second line appears at 6 ns for several SC sectors & TOF Δt peaks seems significantly different than the ideal plot
30326----->30274 TOF Δt peaks seems significantly different than the ideal plot


Recon. SC 1

Run Number comments
30786,30771 ~10 times less statistics
30476 An order of magnitude less statistics& SC q+ and q- (β Vs P are bad)
30337/6/5/4/3 ~ 2 to 3 times less statistics


Recon. SC 2

Run Number comments
30786,30771 ~10 times less statistics
30476 An order of magnitude less statistics
30337/6/5/4/3 ~ 2 to 4 times less statistics
30300----->30274 At least 3 times more accidentals


Recon. TOF 1

Run Number comments
30786,30771 ~10 times less statistics
30476 An order of magnitude less statistics
30337 ~ 2 to 4 times less statistics & proton band is slightly higher than the expected line
30336/5/4/3 ~ 2 to 4 times less statistics
30298,30274 0.05 Shift in beta for the charged tracks


Recon. TOF 2

Run Number comments
30786,30771 7 to 10 times less statistics
30476 An order of magnitude less statistics & shifted Δt(TOF – RF)
30337 ~ 2 to 4 times less statistics & shifted Δt(TOF – RF)
30336/5/4/3 ~ 2 to 4 times less statistics
30274 More than 1 ns shift in Δt(TOF – RF)


Thomas

Recon. Tracking 1

Run Number comments
30786 ~10 times less statistics


Everything else looks to be within statistical fluctuations or is already flagged "not good".

Recon. Tracking 2

Run Number comments
30786 ~10 times less statistics


Everything else looks to be within statistical fluctuations or is already flagged "not good".

Recon. Tracking 3

Run Number comments
30786 ~10 times less statistics


Everything else looks to be within statistical fluctuations (< 1keV) or is already flagged "not good".

Ashley

I compared the track/hit matching histograms using run 30323 as a reference. The changes from the expected plots are as follows:

Recon. BCAL Matching, Ver15


Run Number comments
31049 low statistics




Recon. FCAL Matching, Ver15


Run Number comments
31049 low statistics
most runs Track distance vs θ - fluctuation at 14 deg.
30980-30982 Structure in FCAL Match Rate (p>1 GeV/c) plot



Recon. TOF Matching, Ver15

Run Number comments
30770, 30480 low statistics?



TOF Matching, Ver15

Run Number comments
30274-30477, 30481-30769, 30778-30966, 30992-31036 low statistics in projected horizontal paddle 19


Monitoring Checks

Section Person
BCAL/FCAL Timing
ST/TOF Timing Mahmoud
Tagger Timing
BCAL Reconstruction
FCAL Reconstruction
ST/TOF Reconstruction Mahmoud?
Tracking Reconstruction (CDC) Thomas
Tracking Reconstruction (FDC) Alex A.
Track/hit Matching Ashley
rho Analysis Alex A.
omega Analysis

Notes


Run Is Good? Notes
30218 x  ?
30220 x  ?
30262 x  ?
30263 x  ?
30264 x  ?
30265 x  ?
30266 x  ?
30268 x  ?
30269 x  ?
30270 x  ?
30271 x  ?
30273 x  ?
30274 AMO, detector checkout - run begins!
30275 AMO, detector checkout, long
30276 AMO, detector checkout
30277 AMO, detector checkout
30278 x PERP, TAGM calibration problems
30279 PARA
30280 PERP
30281 PERP
30282 PARA
30283 PARA
30284 PARA-135
30285 PARA-135
30286 PERP-45
30298 PERP-45
30299 PARA-135
30300 PERP-45
30310 x trigger studies?
30314 x AMO, short run, beam tripping
30317 x trigger studies?
30319 x trigger studies?
30320 AMO
30321 PARA
30322 PARA
30323 PARA
30324 PERP
30326 PERP
30327 PARA
30329 PARA
30330 PERP
30331 PARA-135
30332 PERP-45
30333 x PERP-45, empty target
30334 x PARA, empty target
30335 x PARA, empty target
30336 x PERP, empty target
30337 x empty target
30342 x DAQ studies
30343 PARA
30344 PERP
30345 PARA
30346 PERP
30347 PARA-135
30348 PERP-45
30349 PARA-135
30350 PERP-45
30351 AMO
30352 AMP
30353 x PARA, junk
30355 PARA
30356 x trigger studies
30357 x trigger studies
30359 x trigger studies
30360 x trigger studies
30361 PERP
30380 PARA
30381 PERP
30383 PERP
30384 PARA-135
30385 PARA-135
30386 PERP-45
30387 PERP-45
30388 AMO
30389 AMO
30390 AMO
30393 x trigger studies
30394 x trigger studies
30395 x trigger studies
30396 x trigger studies
30397 x trigger studies
30400 x high intensity, crash
30401 PERP-45
30402 PERP-45
30403 PARA-135
30404 PARA-135
30405 AMO, high trigger rate ~50kHz, updated DAC calibration for FCAL and BCAL
30406 AMO
30407 AMO
30408 PARA
30409 PERP
30410 PERP
30411 PARA-135
30420 PARA-135
30421 PERP-45
30422 PARA
30423 x PARA, low live time, let's junk it
30424 PARA
30427 x junk
30428 PARA
30429 PARA
30430 x PERP
30431 PERP
30432 PERP
30433 PARA-135
30434 PERP-45
30436 AMO
30437 AMO
30441 PARA
30442 PERP
30446 PERP
30447 PERP
30448 PARA-135
30449 PERP-45
30450 PERP
30451 PARA
30452 PARA-135
30453 PARA-135
30454 PERP-45
30455 AMO
30456 x AMO, DAQ problems
30457 x AMO, DAQ problems
30458 x AMO, DAQ problems
30459 AMO
30460 PERP
30461 PARA
30462 PARA-135
30463 PARA-135
30464 PERP-45
30465 PERP
30466 PARA
30467 PARA-135
30468 PARA-135
30469 PERP-45
30470 PERP-45
30471 AMO
30473 PERP
30474 PARA
30476 x PARA-135, short run
30477 PARA-135
30479 x PARA-135, short run
30480 PARA-135
30481 PERP-45
30482 PERP-45
30483 x short, throw?
30484 PERP-45
30485 PERP
30486 PARA
30487 PARA-135
30488 PERP-45
30489 AMO
30490 PERP
30491 x PERP, junk
30492 x bad data?
30493 PERP
30494 PARA
30495 PARA-135
30496 PARA-135
30497 PERP-45
30499 PERP
30561 x TRD
30562 x TRD
30563 x TRD
30564 x empty target
30565 x target filling
30566 PERP, long mode
30567 PERP
30568 PARA
30570 PARA-135
30571 PERP-45
30573 x bad DAQ
30575 AMO
30577 PERP
30578 PERP
30579 PARA
30580 PARA
30581 PARA-135
30582 PARA-135
30583 x bad DAQ
30584 x bad DAQ
30585 PERP-45
30586 PERP
30587 PARA
30588 x bad DAQ
30589 PARA-135
30590 PARA-135
30591 PERP-45
30592 PERP-45
30593 AMO
30595 PERP
30596 PERP
30597 PARA
30598 PARA
30599 x DAQ crash
30600 PARA-135
30601 x DAQ crash
30602 PARA-135
30607 PARA-135
30608 PERP-45
30610 PERP
30611 PARA
30612 PARA-135
30614 PERP-45
30615 x cosmic
30616 PERP-45
30618 AMO
30620 PERP
30621 PARA
30622 PARA
30623 PARA-135
30624 PERP-45
30625 PERP
30626 PERP
30627 PARA
30628 x DAQ trouble
30629 PARA-135 some beam trips
30630 PARA-135
30631 x DAQ trouble
30632 PERP-45
30633 PERP-45
30634 AMO
30635 AMO
30636 PERP
30637 PARA
30638 PARA-135
30639 PERP-45
30641 PERP-45
30642 PERP-45
30643 PERP-45
30647 x PERP, includes TRD data (livetime reduced)
30648 PERP
30649 PARA
30650 PARA-135
30651 PERP-45
30652 AMO
30653 PERP
30654 PARA, quality?
30655 PARA-135, quality?
30656 PERP-45
30657 PERP
30658 PARA
30659 PARA-135
30660 PERP-45
30666 PERP-45
30667 AMO
30668 PERP
30670 x trigger studies?
30671 x trigger studies?
30672 PARA
30673 PARA-135
30674 PERP-45
30675 PERP-45
30676 PERP
30677 PARA
30678 PARA-135
30679 PARA-135
30680 PERP-45
30682 AMO
30684 PERP
30685 x junk
30686 PARA
30687 PARA-135 ?
30688 PARA-135
30690 PERP-45
30693 PERP
30694 PARA
30695 PARA-135
30696 PARA-135
30697 PERP-45
30698 AMO
30699 AMO
30701 PERP
30728 x empty target
30729 x filling target
30730 PERP
30731 PARA
30732 PARA-135
30733 PARA-135
30734 PERP-45
30735 PERP
30736 PARA
30737 PARA-135
30738 PERP-45
30739 AMO
30740 PERP
30741 PERP
30742 PARA
30743 PARA
30744 PARA-135
30745 PARA-135
30748 x junk?
30749 PERP-45
30750 x PERP-45, beam steering problems
30751 x PERP-45, junk
30754 PERP-45
30760 x trigger test
30763 x trigger test
30764 x trigger test
30765 x trigger test
30769 AMO
30770 PERP
30771 x TAGM test
30772 x TAGM test
30773 x TAGM test
30774 x TAGM test
30775 x TAGM test
30776 x TAGM test
30778 PERP
30779 PARA
30780 PARA
30783 PARA-135
30784 PARA-135
30785 PERP-45
30786 PERP-45, long mode
30787 PERP-45
30788 AMO
30792 x PERP, tests
30794 x PERP, tests
30796 PERP, start of 50kHz running!
30797 PERP
30798 x PERP, junk
30800 PERP
30801 PARA
30802 PARA
30803 PARA-135
30804 PARA-135
30805 PERP-45
30806 x PERP-45, short run
30807 PERP-45
30808 AMO
30809 PERP
30810 PERP
30811 PERP
30812 PARA
30813 PARA
30815 PARA-135
30816 PARA-135
30817 x PERP-45, junk
30818 PERP-45
30819 x PERP-45, short run
30820 x PERP-45, DAQ problems
30821 PERP-45, livetime problems
30822 AMO
30823 AMO
30824 PERP
30825 x PERP, livetime problems
30826 PERP
30827 PERP
30829 PARA
30830 PARA
30832 x short, bad
30833 PARA
30834 PARA-135
30835 PARA-135
30836 PERP-45
30837 x PERP-45, ~50% livetime
30838 PERP-45
30839 AMO
30840 PERP
30841 PERP
30842 PERP, maybe bad?
30843 PARA
30844 PARA
30847 PARA-135
30848 PARA-135
30849 x PERP-45, setup for TAC run
30855 PERP-45
30856 PERP-45
30857 AMO
30858 PERP
30859 PERP
30873 x trigger problems
30874 x trigger problems
30883 x PERP, noisy FCAL, FCAL red light
30884 x AMO, noisy FCAL, FCAL red light
30886 x AMO, noisy FCAL
30887 x AMO, noisy FCAL
30888 AMO
30889 PERP
30890 PARA
30891 PERP
30893 PARA
30894 PARA, long mode
30895 PARA-135
30896 PARA-135
30897 x bad
30898 PERP-45
30899 PERP-45
30900 AMO
30902 PERP
30903 PERP
30907 x trigger studies
30920 PERP
30921 x bad DAQ
30922 x problem config, stopped early
30923 PARA
30924 PERP
30926 PARA-135
30927 PARA-135
30928 PERP-45
30929 PERP-45
30930 AMO
30931 x short, likely bad
30939 x fixing problems
30940 x short, likely bad
30941 x AMO, trigger tests
30942 x AMO, trigger tests
30943 x 0/90, various tests
30944 x PERP, various tests
30946 x PERP, various tests
30947 PERP, run with fADC250 compression
30948 x short, likely bad?
30949 x PERP, empty target run
30951 PERP
30952 PERP
30953 PARA
30954 PARA
30955 PARA-135
30956 PARA-135
30957 PERP-45
30958 PERP-45
30959 AMO
30961 PERP
30962 PERP
30963 PARA
30964 PARA
30965 PARA-135
30966 PARA-135
30967 x PERP-45, potential trigger problems, where one of the FCAL boards died
30969 x tests
30974 x tests
30975 x PERP-45, FCAL hole
30977 x junk
30978 x beam tripping
30980 AMO, low rate, 2e-5, FCAL hole
30981 AMO, low rate, 2e-5, FCAL hole
30982 AMO, low rate, 2e-5, FCAL hole
30988 x testing
30990 x testing
30992 AMO
30993 PERP-45
30994 PERP-45
30995 PERP
30996 PERP
30997 x PERP, junk
30998 PARA
30999 PARA
31000 PARA
31001 PARA-135
31002 PARA-135
31003 PERP-45
31004 PERP-45
31005 AMO (is it good?)
31009 x DAQ debugging
31011 x DAQ debugging
31012 x DAQ debugging
31015 x fdc threshold tuning
31018 AMO
31022 x junk
31023 AMO
31029 PERP
31031 PARA
31032 PARA-135
31034 PERP-45
31036 AMO
31037 x AMO, FDC HV @ 2225V
31046 low lumi run, PERP, 50 nA
31047 x test of min-bias trigger, 2e-5 AMO @ 100 nA
31049 low lumi run, AMO 2x10^-5, 200 nA
31050 low lumi run, AMO 1x10^-4, 100 nA
31051 low lumi run, AMO 1x10^-4, 100 nA
31053 low lumi run, AMO 1x10^-4, 100 nA
31054 low lumi run, AMO 1x10^-4, 100 nA
31055 low lumi run, AMO 1x10^-4, 100 nA
31056 low lumi run, AMO 1x10^-4, 100 nA
31057 low lumi run, AMO 1x10^-4, 100 nA

Edits

  • 30000 - 30480 --- sdobbs
  • 30481 - 30621 --- sdobbs
  • 30622 - 30653 --- njarvis, haven't run is_production rcdb tagging script yet
  • 30654 - 30829 --- sdobbs
  • 30830 - 31057 --- sdobbs