# Issues with wrapping
 
# Issues with wrapping
#* small damages (holes) of the Aluminized mylar, See Fig. 2. Can be fixed using Al sticky tape, thickness < 20 um
+
#* small damages (holes) in the Aluminized mylar, see Fig. 2. This can be fixed by using Al sticky tape, thickness < 20 um
#* Al layer in the wrapping foil peels off or adheres to the glass block (penetrates into the block):  
+
#* Aluminum layer in the wrapping foil peels off or adheres to the glass block (penetrates into the block):  
#** Al layer peeled off resulting in some transparent holes in the wrapping.
+
#** Fig. 3 and Fig. 4:  Aluminum layer peeled off resulting in some transparent holes in wrapping.
 +
#** Recommend to disassemble and re-wrap the lead glass block

Revision as of 11:02, 18 October 2022


Possible issues with lead glass module

  1. Broken lead glass, see Fig. 1 (module taken from the FCAL prototype)
    • some problems can be seen by looking on the lead glass block from the side opposite to the PMT
    • lead glass module has to be disassembled; the glass block has to be replaced.
  2. Issues with wrapping
    • small damages (holes) in the Aluminized mylar, see Fig. 2. This can be fixed by using Al sticky tape, thickness < 20 um
    • Aluminum layer in the wrapping foil peels off or adheres to the glass block (penetrates into the block):
      • Fig. 3 and Fig. 4: Aluminum layer peeled off resulting in some transparent holes in wrapping.
      • Recommend to disassemble and re-wrap the lead glass block