Difference between revisions of "Visual inspection"

From GlueXWiki
Jump to: navigation, search
(Possible issued with lead glass module)
(Possible issues with lead glass module)
Line 21: Line 21:
 
# Broken lead glass, see Fig. 1 (module taken from the FCAL prototype)
 
# Broken lead glass, see Fig. 1 (module taken from the FCAL prototype)
 
#*  some problems can be seen by looking on the lead glass block from the side opposite to the PMT
 
#*  some problems can be seen by looking on the lead glass block from the side opposite to the PMT
#*  lead glass module has to be disassembled
+
#*  lead glass module has to be disassembled; the glass block has to be replaced.
 
# Issues with wrapping
 
# Issues with wrapping
 
#* small damages (holes) of the Aluminized mylar, See Fig. 2. Can be fixed using Al sticky tape, thickness < 20 um
 
#* small damages (holes) of the Aluminized mylar, See Fig. 2. Can be fixed using Al sticky tape, thickness < 20 um
 
#* Al layer in the wrapping foil peels off or adheres to the glass block (penetrates into the block):  
 
#* Al layer in the wrapping foil peels off or adheres to the glass block (penetrates into the block):  
 
#** Al layer peeled off resulting in some transparent holes in the wrapping.
 
#** Al layer peeled off resulting in some transparent holes in the wrapping.

Revision as of 10:56, 18 October 2022


Possible issues with lead glass module

  1. Broken lead glass, see Fig. 1 (module taken from the FCAL prototype)
    • some problems can be seen by looking on the lead glass block from the side opposite to the PMT
    • lead glass module has to be disassembled; the glass block has to be replaced.
  2. Issues with wrapping
    • small damages (holes) of the Aluminized mylar, See Fig. 2. Can be fixed using Al sticky tape, thickness < 20 um
    • Al layer in the wrapping foil peels off or adheres to the glass block (penetrates into the block):
      • Al layer peeled off resulting in some transparent holes in the wrapping.