<gallery>
 
<gallery>
Image:glass_wrap_prob.jpg | Fig. 2 Wrapping issue: teared Aluminized mylar, fix without disassembling module
+
Image:glass_wrap_prob.jpg | Fig. 2 Wrapping issue: teared Aluminized mylar, fix without disassembling module
Image:glass_glass_prob.jpg | Fig. 4 Broken glass
+
Image:glass_wrap_prob1.jpg | Fig. 3 Wrapping issue: teared Aluminized mylar, fix without disassembling module
image:glass_color_prob.jpg | Fig. 1 Colored lead glass module, possible radiation damage issue. Can be seen without removing steel flange.
+
Image:glass_wrap_prob2.jpg | Fig. 4 Wrapping issue: teared Aluminized mylar, fix without disassembling module
Image:glass_color_prob1.jpg | Fig. 2 Colored lead glass module
+
Image:glass_pmt_prob.jpg | Fig. 5 Damaged PMT
+
 
</gallery>
 
</gallery>
  
 +
<gallery>
 +
Image:glass_pmt_prob.jpg | Fig. 5 Damaged PMT
 +
Image:glass_color_prob.jpg | Fig. 6 Colored lead glass module, possible radiation damage issue. Can be seen without removing steel flange.
 +
Image:glass_color_prob1.jpg | Fig. 7 Colored lead glass module 
 +
</gallery>
  
 
== Possible issued with lead glass module ==
 
== Possible issued with lead glass module ==

Revision as of 10:47, 18 October 2022

Possible issued with lead glass module

  1. Broken lead glass, see Fig. 1 (module taken from the FCAL prototype)
    • some problems can be seen when looked on the lead glass block from the side opposite to the PMT
    • lead glass module has to be disassembled
  2. Issues with wrapping
    • small damages (holes) of the Aluminized mylar, See Fig. 2. Can be fixed using Al sticky tape, thickness < 20 um
    • Problems with the Al layer in the wrapping foil:
      • Al layer peeled off resulting in some transparent holes in the wrapping.