Difference between revisions of "Tracking-11-10-2022"

From GlueXWiki
Jump to: navigation, search
(Created page with "# [https://jlab-org.zoomgov.com/j/1619020033?pwd=NlRKeU9lSzhFc3FIVE5hS1dBOTF3dz09 meeting link to zoomgov] # [https://halldweb.jlab.org/wiki/index.php/FDC_Meetings meeting pag...")
 
(Minutes)
Line 12: Line 12:
  
 
= Minutes =
 
= Minutes =
 +
#* Lubomir found some problems with FDC preamps following the recent raw data run 111255 - the first for this run period. It looks like the gain of 3 preamps has dropped to half, and there are a handful of other isolated channels which look like their signal height readout is missing bits. 
 +
See https://logbooks.jlab.org/entry/4081020
 +
#* We briefly discussed Sasha's inquiry about removing the tungsten plate for beam systematics tests.  This should not pose a problem to the drift chambers at 200nA beam current.  However if this goes ahead we should make sure that RoboCDC is switched OFF, as the HVB currents in the CDC would double.  The logbook entry from 19 April 2018 about the W plate insertion is at https://logbooks.jlab.org/entry/3563927  't w
 +
#* We discussed plans for maintenance during the new year break.  We do not propose to touch the CDC electronics at present. Since another set of channels developed the same kind of problem as I-1 and I-26, where they trip and have to be coaxed very gradually back up to full HV, and these were solved by switching to a different HV module, the leading theory is now that there is some problem with the HV module and probably not with accumulation of debris on the HV boards.  Maybe it is overheating.

Revision as of 11:51, 16 November 2022

  1. meeting link to zoomgov
  2. meeting page

Thu 11/10/2022 2pm

Agenda

  1. Detector status:
    • CDC
    • FDC
  2. Tracking
  3. GEM-TRD
  4. Other

Minutes

    • Lubomir found some problems with FDC preamps following the recent raw data run 111255 - the first for this run period. It looks like the gain of 3 preamps has dropped to half, and there are a handful of other isolated channels which look like their signal height readout is missing bits.

See https://logbooks.jlab.org/entry/4081020

    • We briefly discussed Sasha's inquiry about removing the tungsten plate for beam systematics tests. This should not pose a problem to the drift chambers at 200nA beam current. However if this goes ahead we should make sure that RoboCDC is switched OFF, as the HVB currents in the CDC would double. The logbook entry from 19 April 2018 about the W plate insertion is at https://logbooks.jlab.org/entry/3563927 't w
    • We discussed plans for maintenance during the new year break. We do not propose to touch the CDC electronics at present. Since another set of channels developed the same kind of problem as I-1 and I-26, where they trip and have to be coaxed very gradually back up to full HV, and these were solved by switching to a different HV module, the leading theory is now that there is some problem with the HV module and probably not with accumulation of debris on the HV boards. Maybe it is overheating.