Sep 21, 2010 Calibration/Monitoring

From GlueXWiki
Revision as of 11:40, 22 September 2010 by Zisis (Talk | contribs)

Jump to: navigation, search

Teleconference Time: 07:00 CST (Regina) /14:00 GMT+2 (Athens)

  • Direct Polycom point-to-point connection

Items for followup from previous meeting(s)

  1. None

Documents to Review

Tentative Agenda

  1. Winston cone versus Trapezoidal lightguides
  2. FCAL block plus Plexiglas test
  3. Review LED monitoring system

Minutes

Attendees: Christina, George V., Pavlos (Athens), Zisis (Regina)

  1. Winston cone versus Trapezoidal lightguides: the current setup does not seem to be very robust: it includes a LED plus diffuser on one end of the baby cal and a PMT coupled on the other end via either lightguide using optical grease. Initial tests showed that trap/wico=0.8. A horizontal scan showed that max(trap)=max(wico), at different x positions. The setup needs to be improved, by ensuring that the diffuser system produces more uniform light.
  2. FCAL block plus Plexiglas test: these are in progress, and a horizontal position scan has been measured.
  3. Review LED monitoring system