Difference between revisions of "Sep 21, 2010 Calibration/Monitoring"
From GlueXWiki
(→Tentative Agenda) |
|||
Line 13: | Line 13: | ||
# FCAL block plus Plexiglas test | # FCAL block plus Plexiglas test | ||
# Review LED monitoring system | # Review LED monitoring system | ||
+ | |||
+ | = Minutes = | ||
+ | |||
+ | ''Attendees: Christina, George V., Pavlos (Athens), Zisis (Regina)'' | ||
+ | |||
+ | # Winston cone versus Trapezoidal lightguides: the current setup does not seem to be very robust: it includes a LED plus diffuser on one end of the baby cal and a PMT coupled on the other end via either lightguide using optical grease. Initial tests showed that trap/wico=0.8. A horizontal scan showed that max(trap)=max(wico), at different x positions. The setup needs to be improved, by ensuring that the diffuser system produces more uniform light. | ||
+ | # FCAL block plus Plexiglas test: these are in progress, and a horizontal position scan has been measured. | ||
+ | # Review LED monitoring system | ||
+ | ## |
Revision as of 11:40, 22 September 2010
Teleconference Time: 07:00 CST (Regina) /14:00 GMT+2 (Athens)
- Direct Polycom point-to-point connection
Contents
Items for followup from previous meeting(s)
- None
Documents to Review
Tentative Agenda
- Winston cone versus Trapezoidal lightguides
- FCAL block plus Plexiglas test
- Review LED monitoring system
Minutes
Attendees: Christina, George V., Pavlos (Athens), Zisis (Regina)
- Winston cone versus Trapezoidal lightguides: the current setup does not seem to be very robust: it includes a LED plus diffuser on one end of the baby cal and a PMT coupled on the other end via either lightguide using optical grease. Initial tests showed that trap/wico=0.8. A horizontal scan showed that max(trap)=max(wico), at different x positions. The setup needs to be improved, by ensuring that the diffuser system produces more uniform light.
- FCAL block plus Plexiglas test: these are in progress, and a horizontal position scan has been measured.
- Review LED monitoring system