Difference between revisions of "PS instrumentation studies"

From GlueXWiki
Jump to: navigation, search
Line 1: Line 1:
 
# Decided what magnet to use: 2 m long 30D72 magnet with the maximum field of 2 T.
 
# Decided what magnet to use: 2 m long 30D72 magnet with the maximum field of 2 T.
# The vaccum chamber design has to be optimized. Current chamber length 1.5 m.
+
# The vaccum chamber design has to be optimized. The current length of the VC is 1.5 m.
 
# Consider symmetric detectors covering the energy range from 3 GeV to 6 GeV.
 
# Consider symmetric detectors covering the energy range from 3 GeV to 6 GeV.
  
Line 16: Line 16:
  
 
[[Image:ps_mult_scat.jpg |350px]]
 
[[Image:ps_mult_scat.jpg |350px]]
 +
 +
# Impact of the beam profile distribution and beam motion on the energy resolution
 +
 +
 +
# Counter Rates

Revision as of 09:55, 3 May 2010

  1. Decided what magnet to use: 2 m long 30D72 magnet with the maximum field of 2 T.
  2. The vaccum chamber design has to be optimized. The current length of the VC is 1.5 m.
  3. Consider symmetric detectors covering the energy range from 3 GeV to 6 GeV.

Detector width (position from the beamline) as a function of the magnetic field:

Det width.jpg For the magnetic field of 1.5 T the detector width should be about 30 cm.


Contribution to the position/energy resolution from multiple scattering. The converter is placed 1 m in front of the magnet. B = 1 T.

(the energy resolution of the scintillator counters corresponding to the default design is 12 MeV for 24 counters and 17 MeV for 8 counters).

Ps mult scat.jpg

  1. Impact of the beam profile distribution and beam motion on the energy resolution


  1. Counter Rates