== Gas issues ==
 
== Gas issues ==
  
- CO2 was monitored for two weeks. Two sets of measurements are shown in the above plots. First plot: starting with certified gas (60% CO2) in EEL126 then moving the sensor (without disconnecting from the laptop) to the Hall and connecting
+
- CO2 was monitored for two weeks. Two sets of measurements are shown in the above plots. First plot: starting with certified gas (60% CO2) in EEL126 then moving the sensor (without disconnecting from the laptop) to the Hall and sensing first FDC1 for several days and at the end switching the solenoid valve to the CDC. Shown are the atmospheric pressure, sensor results and the ones corrected for the pressure: ~(dleta_p/p)^n, where the power n was chosen, on the first plot, to make the calibration measurements flat (n=2.4). Second plot is the same except n=1.4 which makes the CDC measurements plat. Third and fourth plots represent the second set of measurements, starting with 5 days sensing CDC and then few days FDC1.
 +
 
 +
- Discussions: 
 +
 
 