January 19, 2011 Tracking CDC/FDC

From GlueXWiki
Revision as of 11:26, 20 January 2011 by Nsjarvis (Talk | contribs) (CDC Update)

Jump to: navigation, search

Meeting Time and Place

Wednesday January 19, 2011 at 11:00am At Jefferson Lab, the meeting will be held in F326

Connections

To connect from the outside, please use ESNET
1.) ESNET: 8542553
2.) Phone:
+1-800-377-8846 : US
+1-888-276-7715 : Canada
+1-302-709-8424 : International
then enter participant code: 77438230# (remember the "#")

Agenda

  1. Announcements
  2. The Collaboration Meeting
  3. FDC Update logbook
  4. CDC Update [1]
  5. Tracking Issues

Minutes

Participants: Curtis, Naomi, Beni, Simon, Lubomir, Dave and Mark

Collaboration Meeting

Curtis has scheduled one talk each for the FDC and CDC groups. The next section is Offline Reconstruction which will follow on nicely, there will probably be an overview by Mark and reports on Dmitri's calibration database and tracking status. Beni and Lubomir will let Curtis know in the next few days if they need to make another talk on the FDC.

FDC Update

CDC Update

Curtis reported on the CDC construction status: the remaining straws in row 10 will be put into place today and glued tomorrow. We are completing about one layer each week. An undergrad student is working on cutting more straws to size and others are checking the parts - there are still a few 1000 pinholders to be checked and cleaned.

Installing the straws takes 3 people, usually these are the 2 techs and Gary; Curtis and Naomi have been trained for use if needed.

We have not found any more bad resistance straws - these all came from one section of the box. Initially found 2 straws where where the straw-donut glue-joint made no electrical connection, this has not been seen since then. The resistance of each straw and straw-donut is being checked.

Gary has made some modifications to Slava's test electrical hookup. He found that it can be sealed with less torque than initially specified. He also found that the wires leaked gas (along the mesh) but can sealed with cyanoacrylic over the rubber boot at the pin end.

We plan to train students to prepare the wires - cut, strip, solder ball, seal, add the rubber boot and make sure that the finished wires have low resistance.

Cleanroom... our particulate meter was returned from servicing and indicates that we still have a class 2000 cleanroom. The low humidity has been causing static electricity problems (eg giving people a small shock when they touch the table) so we have been trying to raise humidity and also grounded the cleanroom and the table to the building ground. After installing the straws in the CDC, the ends are sealed with a plastic cap to prevent dust from collecting inside them.

Naomi has been working installing Xilinx software so that she can download software from Gerard into the new fADC to increase the pre-trigger buffer time. She is preparing to send some of our MFCs back to MKS for recalibration.

Tracking Issues