Present: Dave Meekins, Jim Fochtman, Mark Ito, Chris Keith, Tim Whitlatch,

Review Slides

Jim's Presentation

Notes

  • Target volume of 0.4 l of liquid hydrogen expands to about 1000 l at STP.
  • Hall B pumps on scattering chamber with turbo pump all the time.
  • Jim added Lakeshore Cernox temperature sensors at various heights.
    • Chris requested that they be removable, to replace defective ones.
  • Chris recommends the Lakeshore Model 218 8-channel temperature monitor read-out unit.
  • System should be able to liquefy helium-4, taking it down to 3 K.
  • Jim will look into adding a relief tube to by-pass normal feed in case of boil off of target liquid.