New HVB
Max amplitude histograms for
cosmics at 2100V with 50/50 Ar/CO2
Left: new preamp and new HVB, ADC half-range; right: old preamp and HVB, ADC full-range
Purple: untracked, blue: tracked (less of this with old preamp because of 2 dead central channels)
Cosmics with premixed 50/50 Ar/CO2, New preamp and new HVB, ADC half-range
NB expanded x-axis, zoomed to 1000ch from 4096.
Left: 2000V; Right: 2050V; Far Right: 1950V (too low for software threshold of ped_mean+5sigma)
Fe55 with premixed 50/50 Ar/CO2, New preamp and new HVB, ADC half-range, for a range of HV
Left: max amplitude histogram; Right: energy
Mean pedestal is around 455ch, this has been subtracted from the amplitude histogram so at 2200V it is the fADC saturating rather than the preamp.
The beginnings of a new HV efficiency plot
Scope traces for premixed 50/50 Ar/CO2
older pics below (trigger was too high)