CDC tilted detectors
From GlueXWiki
Prototype at 2100V, 50/50 Ar/CO2
The paler lines are fits to a Landau dist, the darker lines are fits to a Landau-Gaussian convolution.
Ratio of MPVs of Landau fit to integration histograms... horizontal : 30deg : 45deg is 5110 : 6620 : 7160 or 1 : 1.29 : 1.40 Ideally this should be 1 : 1.15 : 1.41
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
218 | 229 | 256 | 264 | 264 | 285 | 1.17 | 1.15 | 1.21 | 1.24 |
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
5167 | 5301 | 6722 | 7446 | 7163 | 8113 | 1.30 | 1.40 | 1.39 | 1.53 |
50/50 premix gas, unfiltered, label reads analysis 49.1 CO2, so this is 51/49 Ar/CO2
2100V with prototype horizontal, scintillators vertical
Estimated max drift time is 650ns
Repeating the tilted measurements just in case the scintillators moved slightly between measurements
The paler lines are fits to a Landau dist, the darker lines are fits to a Landau-Gaussian convolution.
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
394 | 419 | 378 | 413 |
Horizontal MPV | 30 degrees MPV | 45 degrees MPV | 30 degrees:horizontal | 45 degrees:horizontal | |||||
---|---|---|---|---|---|---|---|---|---|
Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss | Landau | Landau-Gauss |
9049 | 10413 | 9291 | 10591 |