* We would like to situate the best chambers downstream. Looking at the efficiency curves in Figs 11 and 12 of [https://halldweb.jlab.org/DocDB/0053/005373/002/MWPC_production_testing.pdf GlueX-doc-5373] I made a table of two parameters: the inefficiency at th=0, and the threshold where the efficiency starts climbing. The best chambers have lower inefficiency and higher inflection threshold. Based on this criteria, I suggest the following ordering: | * We would like to situate the best chambers downstream. Looking at the efficiency curves in Figs 11 and 12 of [https://halldweb.jlab.org/DocDB/0053/005373/002/MWPC_production_testing.pdf GlueX-doc-5373] I made a table of two parameters: the inefficiency at th=0, and the threshold where the efficiency starts climbing. The best chambers have lower inefficiency and higher inflection threshold. Based on this criteria, I suggest the following ordering: | ||
Line 22: | Line 21: | ||
− | * | + | * We have left all chambers in alphabetical order, except for Haldor that is slightly worse than Denethor an Eowyn. Note that Haldor is almost as good as Denethor, so keeping complete alphabetical order is almost as good as the list above. The next table has the parameters for the two spares. |
{| class="wikitable" | {| class="wikitable" |
Revision as of 16:59, 31 January 2022
- We would like to situate the best chambers downstream. Looking at the efficiency curves in Figs 11 and 12 of GlueX-doc-5373 I made a table of two parameters: the inefficiency at th=0, and the threshold where the efficiency starts climbing. The best chambers have lower inefficiency and higher inflection threshold. Based on this criteria, I suggest the following ordering:
Number | MWPC | inefficiency (x10^-3) | Plateau width
(x10^6 integrated counts) |
---|---|---|---|
1 | Arwen | 3 | 6 |
2 | Bilbo | 1.8 | 9 |
3 | Celeborn | 1.5 | 6 |
4 | Haldor | 1.2 | 8 |
5 | Denethor | 1.2 | 9 |
6 | Eowyn | 1.3 | 10 |
- We have left all chambers in alphabetical order, except for Haldor that is slightly worse than Denethor an Eowyn. Note that Haldor is almost as good as Denethor, so keeping complete alphabetical order is almost as good as the list above. The next table has the parameters for the two spares.
Number | MWPC | inefficiency (x10^-3) | Plateau width (x10^6 integrated counts) | Comment |
---|---|---|---|---|
7 | Frodo (Th=600) | 11 | 7 | |
8 | Galadriel (Th=500) | 4 | 5 | Inefficient Wire#37 |