Difference between revisions of "CDC signal integration new"
(Created page with "Cosmics, 50/50 gas, 2090V Comparison with old (underpowered pream) data is at CDC_signal_integration Run numbers are 31809 (horiz) and 31810 (45 deg) at 3.0V, 2.09kV {|...") |
|||
Line 4: | Line 4: | ||
− | Run numbers are 31809 (horiz) | + | Run numbers are 31809 (horiz) 31810 (45 deg) and 31811 (30 deg) at 3.0V, 2.09kV |
Line 11: | Line 11: | ||
|+Pedestal height | |+Pedestal height | ||
|[[Image:run_31809_ped20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | |[[Image:run_31809_ped20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | ||
+ | |[[Image:run_31811_ped20.png|thumb|x180px|30 degrees, LV 3.0V, HV 2090V]] | ||
|[[Image:run_31810_ped20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | |[[Image:run_31810_ped20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | ||
|} | |} | ||
Line 17: | Line 18: | ||
|+Pedestal width (std dev of first 100 samples in each event) | |+Pedestal width (std dev of first 100 samples in each event) | ||
|[[Image:run_31809_sig20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | |[[Image:run_31809_sig20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | ||
+ | |[[Image:run_31811_sig20.png|thumb|x180px|30 degrees, LV 3.0V, HV 2090V]] | ||
|[[Image:run_31810_sig20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | |[[Image:run_31810_sig20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | ||
|} | |} | ||
Line 23: | Line 25: | ||
|+Histograms of first max in signal amplitude with hit threshold 4 x pedestal sigma | |+Histograms of first max in signal amplitude with hit threshold 4 x pedestal sigma | ||
|[[Image:run_31809_thr4_maxamp20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | |[[Image:run_31809_thr4_maxamp20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | ||
+ | |[[Image:run_31811_thr4_maxamp20.png|thumb|x180px|30 degrees, LV 3.0V, HV 2090V]] | ||
|[[Image:run_31810_thr4_maxamp20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | |[[Image:run_31810_thr4_maxamp20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | ||
|} | |} | ||
Line 29: | Line 32: | ||
|+Histograms of first max in signal amplitude with hit threshold 5 x pedestal sigma | |+Histograms of first max in signal amplitude with hit threshold 5 x pedestal sigma | ||
|[[Image:run_31809_thr5_maxamp20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | |[[Image:run_31809_thr5_maxamp20.png|thumb|x180px|No tilt, LV 3.0V, HV 2090V]] | ||
+ | |[[Image:run_31811_thr5_maxamp20.png|thumb|x180px|30 degrees, LV 3.0V, HV 2090V]] | ||
|[[Image:run_31810_thr5_maxamp20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | |[[Image:run_31810_thr5_maxamp20.png|thumb|x180px|45 degrees, LV 3.0V, HV 2090V]] | ||
|} | |} |
Revision as of 13:11, 9 July 2012
Cosmics, 50/50 gas, 2090V
Comparison with old (underpowered pream) data is at CDC_signal_integration
Run numbers are 31809 (horiz) 31810 (45 deg) and 31811 (30 deg) at 3.0V, 2.09kV
Data below are for hit threshold 5 x pedestal sigma
Try to calculate dE/dx...
Track length is calculated in the plane parallel to the endplates (no z information) so I scaled it by the tilt of the prototype for the 45 degree data, the upper cosmic trigger scintillator is high up enough that the cosmics are ~ vertical.
The ratio of tilted to horizontal integration means is 1.24 (old data) and 1.28 (new data); it should be 1 (as they are already scaled by track length) so this means the tilted integrations are 25-30% too large.
The ratio of tilted to horizontal MPV of Landau fitted to amplitude/track length histogram is 0.99 (old data) and 1.07 (new data).
The ratio of tilted to horizontal MPV of Landau fitted to (integration to first maximum/tracklength) histogram is 1.07 (old data) and 1.18 (new data).
Look at time over threshold...
Peak duration is # samples after the lower threshold crossing until the ADC value decreases by 2sigma over 2 samples, this is a measure of time over threshold in units of 8ns