Peak length is # samples after the lower threshold crossing until the ADC value decreases by 2sigma over 2 samples, this is a measure of time over threshold in units of 8ns
 
  
  
Line 7: Line 6:
  
  
{| border="0" cellpadding="2"
+
 
|+Peak length vs first maximum in signal amplitude
+
 
|[[Image:run_31757_pklength_maxamp.png|thumb|x250px|No tilt]]
+
|[[Image:run_31765_pklength_maxamp.png|thumb|x250px|Tilted at 45 degrees]]
+
|}
+
  
 
Try to calculate dE/dx
 
Try to calculate dE/dx
Line 35: Line 31:
 
|[[Image:run_31757_intmax1dx.png|thumb|x250px|No tilt]]
 
|[[Image:run_31757_intmax1dx.png|thumb|x250px|No tilt]]
 
|[[Image:run_31765_intmax1dx.png|thumb|x250px|Tilted at 45 degrees]]
 
|[[Image:run_31765_intmax1dx.png|thumb|x250px|Tilted at 45 degrees]]
 +
|}
 +
 +
 +
Look at time over threshold...
 +
 +
Peak duration is # samples after the lower threshold crossing until the ADC value decreases by 2sigma over 2 samples, this is a measure of time over threshold in units of 8ns
 +
 +
 +
{| border="0" cellpadding="2"
 +
|+Peak duration 
 +
|[[Image:run_31757_pklength.png|thumb|x250px|No tilt]]
 +
|[[Image:run_31765_pklength.png|thumb|x250px|Tilted at 45 degrees]]
 +
|}
 +
 +
 +
{| border="0" cellpadding="2"
 +
|+Peak duration / track length
 +
|[[Image:run_31757_pklengthdx.png|thumb|x250px|No tilt]]
 +
|[[Image:run_31765_pklengthdx.png|thumb|x250px|Tilted at 45 degrees]]
 +
|}
 +
 +
 +
 +
{| border="0" cellpadding="2"
 +
|+Peak length vs first maximum in signal amplitude
 +
|[[Image:run_31757_pklength_maxamp.png|thumb|x250px|No tilt]]
 +
|[[Image:run_31765_pklength_maxamp.png|thumb|x250px|Tilted at 45 degrees]]
 
|}
 
|}

Revision as of 15:38, 19 June 2012


50/50 gas, 2100V

        • Caveat: data were with preamp underpowered, 2.65V on preamp card ****



Try to calculate dE/dx

integration from low threshold crossing for next 700ns / track length
No tilt
Tilted at 45 degrees

This does not scale very nicely with track length.


Track length is calculated in the plane parallel to the endplates (no z information) so I scaled it by the tilt of the prototype for the 45 degree data, the upper cosmic trigger scintillator is high up enough that the cosmics are ~ vertical

height of first maximum in signal amplitude / track length
No tilt
Tilted at 45 degrees


signal integrated from low threshold crossing to position of first maximum in signal amplitude / track length
No tilt
Tilted at 45 degrees


Look at time over threshold...

Peak duration is # samples after the lower threshold crossing until the ADC value decreases by 2sigma over 2 samples, this is a measure of time over threshold in units of 8ns


Peak duration
No tilt
Tilted at 45 degrees


Peak duration / track length
No tilt
Tilted at 45 degrees


Peak length vs first maximum in signal amplitude
No tilt
Tilted at 45 degrees