Solenoid dumps

From GlueXWiki
Revision as of 19:45, 9 February 2017 by Gen (Talk | contribs) (Unintended Magnet Dumps)

(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search

Unintended Magnet Dumps

The list is in the reverse chronological order. It does not include very early times.

Date Time Current, A Ramp, A/s PXI file LOG Coil 2 spike? Cause
2017/02/07 13:58:10 1350 0. pxi_20170207_134445.root log middle The refrigerator was accidentally turned off by a crash button
2016/12/14 09:31:50 1150 0.06 pxi_20161214_092531.root log no A positive voltage excursion on Coil 2, mostly 2D (VTT7)
2016/12/13 19:37:15 505 0.2 pxi_20161213_192448.root log before A positive voltage excursion on Coil 2 (≈0.2 s, ≈4 V), mostly 2D (VTT7), followed by a sharp spike on Coil 2D
2016/12/13 17:47:26 740 0.2 pxi_20161213_173728.root log after QD trip by positive voltage excursion 0.2 s on Coil 2, mostly 2D (VTT7)
2016/04/27 18:37:23 1345 0. pxi_20160427_182709.root log no MPS trip
2016/04/08 14:41:08 1200 0. missing log not checked Unknown - a fuse in the MPS?
2016/04/05 13:44:16 1200 0. pxi_20160405_133835.root log after Refrigerator trip
2016/03/28 09:03:29 508 0.2 pxi_20160328_090140.root log after QD was triggered (marginally) by a 2s-long negative voltage excursion on Coil 2. A spike followed 0.15s after the dump.
2016/03/23 12:49:29 800 0. pxi_20160323_124313.root log no lead flow transient
2016/02/12 09:45:59 300 0. pxi_20160212_094433.root log no Slow dump (cause uncertain) triggered QD and fast dump
2016/02/07 17:08:30 890 0.12 pxi_20160207_170730.root log after Slow dump (cause uncertain) triggered QD and fast dump
2016/02/07 14:38:25 1000 0. pxi_20160207_142631.root log no Slow dump (cause uncertain) triggered QD and fast dump