== Drift time spectra ==
 
== Drift time spectra ==
 +
There is a problem with the (minimum) drift time spectra we have for a while. A GARFIELD simulation told us that the maximum drift time is 200 ns. In figure 4.1 on the left side 4 drift time spectra are shown (4 different channels). Two features are not understood:
 +
 +
* The sharp spike in the beginning
 +
* There is no hard cut-off (edge of the straw), instead there is a long tail.
 +
 +
On The right side of figure 4.1 the drift time spectra of the signal coming after the first signal is plotted, the x-axis unit is ns after the first signal. There is a spike close to zero, these are actually not after pulses but probably belong to the first signal. There is also a peak (with a long tail) around 140 ns. This more or less equals 2x the length of the cable between CDC and Shaper board. Those are most likely reflections.
 +
[[Image:drift1.jpg|thumb|none|400px|figure 4.1: Left: drift time spectra of four channels; right: "drift time" spectra of the second signal (x-axis: ns after first signal)]]
 +
 +
 +
All previous data was taken with cosmic rays going through the middle of the straws. Let's see what happens if we move the trigger scintillators to the end (or beginning) of the straws and remake
 +
figure 4.1. This is shown in figure 4.2.
 +
[[Image:drift2.jpg|thumb|none|400px|figure 4.2: Left: drift time spectra of four channels; right: "drift time" spectra of the second signal (x-axis: ns after first signal) - here the cosmic rays went through the straws close to the endplate.]]
 +
 +
There is a difference: there is a hard cut-off - it comes 200 ns too late but it is there. This might be due to a change of the wire position (straw moved or string tension decreased). After measuring the position of the straws it turned out that several tubes were up to 4 mm below where they should be (half the radius). However, several layers appear to have no sag. This is not understood because all tubes seem to show similar behavior. The sagging tubes have been moved up by adding support and we are taking data again.
 +
 +
The spike in the beginning is still not understood + the the fact that the drift time spectra is now twice as long as it should be. So some parts on the to do list are now:
 +
* (double) Check timing of the fADCs  --> the timing is correct -> recheck GARFIELD simulation.
 +
* re-measure wire-tension (The last time this was done was two years ago).
 +
* Perform a tube sag test with a full ring of 1.5 m long tubes

Latest revision as of 11:37, 22 February 2008

Shaper board

As mentioned before there is a problem with three channels on the shaper board - they show a 10x lower gain. Although it is not clear what is causing this Gerard will send a new board with a slightly different gain.

Clock

Kim told us she will send a 125 MHz clock to CMU earliest in the first week of March.

Trigger

In this setup we will are reading out the trigger using one channel of the fADCs. The block pulse coming out of the coincidence unit is attenuated and shaped (against jitter) using a device Gerard gave us. Out of this signal a differential signal is made using another device (also from Gerard). This signal is injected in the shaper board and then read out by the fADC. The real trigger (a NIM signal going to the STOP input on the fADC) is basically the delayed output from the coincidence unit.

A straight line is fit to the rising edge of the trigger signal and the zero crossing (in ns) is dumped into the histogram shown in figure 3.1.

figure 3.1: Histogram of trigger zero crossing time.

One can see too peaks that are 20 ns away from each other. The early peak contains almost 1/3 of the events. For now we do not worry too much because we have other problems.

Drift time spectra

There is a problem with the (minimum) drift time spectra we have for a while. A GARFIELD simulation told us that the maximum drift time is 200 ns. In figure 4.1 on the left side 4 drift time spectra are shown (4 different channels). Two features are not understood:

  • The sharp spike in the beginning
  • There is no hard cut-off (edge of the straw), instead there is a long tail.

On The right side of figure 4.1 the drift time spectra of the signal coming after the first signal is plotted, the x-axis unit is ns after the first signal. There is a spike close to zero, these are actually not after pulses but probably belong to the first signal. There is also a peak (with a long tail) around 140 ns. This more or less equals 2x the length of the cable between CDC and Shaper board. Those are most likely reflections.

figure 4.1: Left: drift time spectra of four channels; right: "drift time" spectra of the second signal (x-axis: ns after first signal)


All previous data was taken with cosmic rays going through the middle of the straws. Let's see what happens if we move the trigger scintillators to the end (or beginning) of the straws and remake figure 4.1. This is shown in figure 4.2.

figure 4.2: Left: drift time spectra of four channels; right: "drift time" spectra of the second signal (x-axis: ns after first signal) - here the cosmic rays went through the straws close to the endplate.

There is a difference: there is a hard cut-off - it comes 200 ns too late but it is there. This might be due to a change of the wire position (straw moved or string tension decreased). After measuring the position of the straws it turned out that several tubes were up to 4 mm below where they should be (half the radius). However, several layers appear to have no sag. This is not understood because all tubes seem to show similar behavior. The sagging tubes have been moved up by adding support and we are taking data again.

The spike in the beginning is still not understood + the the fact that the drift time spectra is now twice as long as it should be. So some parts on the to do list are now:

  • (double) Check timing of the fADCs --> the timing is correct -> recheck GARFIELD simulation.
  • re-measure wire-tension (The last time this was done was two years ago).
  • Perform a tube sag test with a full ring of 1.5 m long tubes