Click the Last Maintenance date to update when the last maintenance was performed
Click the Task to EDIT the long description of this task
Click here to add a new item to the Hall D maintenance maintainer
TASK | Primary Contact | Secondary Contact | PERIOD (days) | NEXT MAINTENANCE | LAST MAINTENANCE |
Vacuum #10 CRYO Tank | rmont@jlab.org | brandsh@jlab.org | 365 | 10/21/25 | 10/21/24 |
Turbopump HiPace 300 Serial # 14950711 Change/inspect oil cartridge 4 years Backing Pump Penta 10 serial # 21587047 Change oil annually (if dirty), wipe down exterior of pumps. Run up pump and record ultimate pressure after 1 hour. Oil is P3 Changed Oil in Backing Pump on 01/12/2016 32,768hrs on Turbopump Operating Fluid Resovoir-# PM 133 302 AU changed out 7/17/17 Backing pump makes a clicking sound when running.7/17/17 2/7/2019Pump checked needs oil changed but not performed due to closing Hall. Will be done at end of run.ASAP. 8/4/21 Backing pump oil changed. 07-12-2023 Backing Pump Oil Changed **Curiously, the hours are the same as they were in 2016, 32768 Hours on the dispaly** 10-21-2024 Backing pump oil change JLab Tag: F2233__ |